|
Various pathophysiologic conditions may be diagnosed or confirmed by examination of the CVP waveform ( Table 32-9 ). One of the most common applications is the rapid diagnosis of cardiac arrhythmias.[311] In atrial fibrillation ( Fig. 32-25 ), the a wave disappears and the c wave becomes more prominent because atrial volume is greater at end-diastole and the onset of systole owing to the absence of effective atrial contraction. Occasionally, atrial fibrillation or flutter waves may be seen in the CVP trace,
Condition | Characteristics |
---|---|
Atrial fibrillation | Loss of a wave |
|
Prominent c wave |
Atrioventricular dissociation | Cannon a wave |
Tricuspid regurgitation | Tall systolic c-v wave |
|
Loss of x descent |
Tricuspid stenosis | Tall a wave |
|
Attenuation of y descent |
Right ventricular ischemia | Tall a and v waves |
|
Steep x and y descents |
|
M or W configuration |
Pericardial constriction | Tall a and v waves |
|
Steep x and y descents |
|
M or W configuration |
Cardiac tamponade | Dominant x descent |
|
Attenuated y descent |
Respiratory variation during spontaneous or positive-pressure ventilation | Measure pressures at end-expiration |
Right-sided valvular heart diseases alter the CVP waveform in different ways.[312] Tricuspid regurgitation ( Fig. 32-26 ) produces abnormal systolic filling of the right atrium through the incompetent valve. A broad, tall systolic c-v wave is inscribed that begins in early systole and obliterates the systolic x descent in atrial pressure. The CVP trace is said to be ventricularized because it resembles right ventricular pressure. Note that this regurgitant wave differs in onset, duration, and magnitude from a normal CVP v wave caused by end-systolic atrial filling from the venae cavae. In patients with tricuspid regurgitation, right ventricular end-diastolic pressure is overestimated by the numeric display on the bedside monitor, which reports a single mean value for CVP. Instead, right ventricular end-diastolic pressure is estimated best by measuring the CVP value at the time of the ECG R wave, before the regurgitant systolic wave (see Fig. 32-26 ). Unlike tricuspid regurgitation, tricuspid stenosis (see Fig. 32-26 ) is a diastolic defect in atrial emptying and ventricular filling. Mean CVP is elevated, and a pressure gradient exists throughout diastole between the right atrium and ventricle. The a wave is unusually prominent and the y descent is attenuated because of the impaired diastolic egress of blood from the atrium. Other conditions that
Figure 32-25
Central venous pressure (CVP) changes caused by cardiac
arrhythmias. A, Atrial fibrillation. Note the absence
of the a wave, a prominent c wave, and a preserved v wave and y descent. This arrhythmia
also causes variation in the electrocardiographic (ECG) R-R interval and left ventricular
stroke volume, which can be seen in the ECG and arterial pressure (ART) traces.
B, Isorhythmic atrioventricular dissociation. In
contrast to the normal end-diastolic a wave in the CVP trace (left
panel), an early systolic cannon wave is inscribed (asterisk,
right panel). The reduced ventricular filling accompanying this arrhythmia
causes decreased arterial blood pressure. C, Ventricular
pacing. Systolic cannon waves are evident in the CVP trace during ventricular pacing
(left panel). Atrioventricular sequential pacing
restores the normal venous waveform and increases arterial blood pressure (right
panel). The ART scale is shown on the left,
the CVP scale on the right. (Redrawn from
Mark JB: Atlas of Cardiovascular Monitoring. New York, Churchill Livingstone, 1998,
Figs. 14-1, 14-5, and 14-16.)
Figure 32-26
Central venous pressure (CVP) changes in tricuspid valve
disease. A, Tricuspid regurgitation increases mean
CVP, and the waveform displays a tall systolic c-v wave that obliterates the x descent.
In this example, the a wave is not seen because of atrial fibrillation. Right ventricular
end-diastolic pressure is estimated best at the time of the electrocardiographic
R wave (arrows) and is lower than mean CVP. B,
Tricuspid stenosis increases mean CVP, the diastolic y descent is attenuated, and
the end-diastolic a wave is prominent. (Redrawn from Mark JB: Atlas of
Cardiovascular Monitoring. New York, Churchill Livingstone, 1998, Figs. 17-3 and
17-15.)
Perhaps the most important application of CVP monitoring is to provide an estimate of the adequacy of circulating blood volume and right ventricular preload. As noted earlier, for this purpose, transmural CVP is always the pressure of physiologic interest. In clinical practice, however, we measure and record pressures referenced to ambient atmospheric pressure. Consequently, accurate interpretation of CVP requires the physician to consider alterations in intrathoracic or juxtacardiac pressure that occur during the respiratory cycle.[294] [303] During spontaneous breathing ( Fig. 32-27 ), inspiration causes a decrease in pleural and juxtacardiac pressure that is transmitted, in part, to the right atrium and lowers CVP. This same decrease in pleural pressure will influence other measured central vascular pressures in similar fashion. Note a subtle, but critically important observation about the measurement of central vascular pressures. Although CVP measured relative to atmospheric pressure decreases during the inspiratory phase of spontaneous ventilation, transmural CVP, the difference between right atrial pressure and juxtacardiac pressure, may actually increase slightly as more blood is drawn into the right atrium. The opposite pattern is observed during positive-pressure ventilation, in which inspiration increases intrathoracic pressure, raises the measured CVP, but decreases transmural CVP because the elevated intrathoracic pressure reduces venous return. In clinical practice, transmural pressures are rarely measured because of difficulty assessing juxtacardiac or intrathoracic pressure.
Figure 32-27
Respiratory influences on the measurement of central
venous pressure (CVP). A, During spontaneous ventilation,
the onset of inspiration (arrows) causes a reduction
in intrathoracic pressure that is transmitted to both the CVP and the pulmonary artery
pressure (PAP) waveforms. CVP should be recorded at end-expiration (mean CVP, 14
mm Hg). B, During positive-pressure ventilation,
the onset of inspiration (arrows) causes an increase
in intrathoracic pressure. CVP is still recorded at end-expiration (mean CVP, 8
mm Hg). (Redrawn from Mark JB: Atlas of Cardiovascular Monitoring. New
York, Churchill Livingstone, 1998, Figs. 16-1 and 16-2.)
Not only can individual CVP waveforms provide unique diagnostic clues about the circulation, but trends in CVP during anesthesia and surgery are also useful in estimating fluid or blood loss and guiding replacement therapy. It is important to remember that the range in normal values is considerable and that small changes in CVP may reflect significant changes in circulating blood volume and right ventricular preload. Additional useful information may be derived from examining how a fluid bolus simultaneously alters CVP and other variables of clinical interest such as blood pressure, urine output, and so forth.
|