Accuracy
All measurements have errors. Error is usually determined by
the comparison of a measurement to a "gold standard" of that measure. Unfortunately,
all measurements, even the so-called gold standards, are subject to errors with respect
to reproducibility. From a clinician's perspective, we can depend on a physiologic
measurement only if it is accurate to the degree required for clinical decision making.
For example, systemic blood pressure can be measured in several ways. We can listen
to Korotkoff sounds by the use of a sphygmomanometer cuff and stethoscope; we can
rely on an oscillometric automated noninvasive blood pressure device, or if we need
continuous measurement, we may place an arterial cannula.[1]
Unfortunately, each of these techniques provides a slightly different arterial blood
pressure value, and each has different sources of error. Our choice of method may
be determined by accuracy or by our needs for the frequency of the data and the ease
of retrieving these data. An automatic oscillometric device is usually chosen over
manual auscultatory measurements for ease of acquisition and reproducibility. Two
people taking auscultatory blood pressure measurements may hear the Korotkoff sounds
at slightly different points and record different blood pressures. The required
accuracy of a clinical monitor is determined by the smallest change in the measured
variable that could affect a clinical decision. The requirements for "absolute accuracy"
(is the measured value correct?) may be different from the requirements for "relative
accuracy" (does the measured value follow trends?).
For example, a pulse oximeter estimates arterial hemoglobin oxygen
saturation by measuring light absorbance. The pulse oximeter saturation estimate,
SpO2
, is compared with the hemoglobin
saturation determined from analysis of an arterial blood sample by a laboratory cooximeter,
SaO2
. Some errors are associated with
arterial blood sampling and in vitro analysis of the sample by the co-oximeter.
Nevertheless, the co-oximeter is considered to be the gold standard in this comparison
of methods.
The pulse oximeter's saturation value can be compared with that
of the co-oximeter by the determination of bias and
precision ( Fig.
30-2
). Bias is the average difference between simultaneous values from
the two methods, or the systematic error. If a pulse oximeter reads an average of
5% higher than the co-oximeter does, it has a bias of 5%, and we can adjust for that
systematic error by recalibrating the device. Precision is the standard deviation
of the difference between the two measurements, and it quantifies the random error,
or "scatter." A higher value for precision indicates a larger random error. (This
statistic may be more appropriately called "imprecision.") If the random error is
too large, the device may not be clinically useful. We can adjust for systematic
error (bias) by recalibration, but there is no way to adjust for random error.[2]