Previous Next

Blood Gas Assessment and Ventilator Management

The therapeutic administration of SpO2 or O2 at 1 ATA settings is usually guided by measurement of arterial blood gases. Because elevation of tissue PO2 is usually the major therapeutic aim with the use of HBO, it would be preferable if blood gases could be used in exactly the same way in this setting. However, it is difficult to attain this goal for several practical reasons. First, blood gas values in a sample taken from a patient in a hyperbaric chamber may not remain stable. In any sample of arterial blood, gas has a tendency to diffuse from blood into the small bubbles that inevitably exist within the syringe unless an extremely careful technique is used to remove them. Under normal clinical conditions at 1 ATA, diffusion of O2 and CO2 from blood into a small bubble in a syringe results in only minor changes in blood gas tension. Changes in partial pressure are buffered by the relatively large content of O2 (because of hemoglobin binding) and CO2 (because of the relatively high solubility of CO2 in plasma). However, in blood samples obtained from patients under hyperbaric conditions, large changes in PO2 may result from delay in measurement of an arterial blood sample, changes that are compounded if the measurement is to be made outside the chamber. In this situation, small bubbles present in the syringe before decompression would be correspondingly increased in size after decompression to 1 ATA. Furthermore, at an ambient pressure of 760 mm Hg, the O2 in the blood sample will be supersaturated and tend to move rapidly out of solution. Finally, it is impossible to calibrate blood gas machines accurately for gas tensions higher than ambient barometric pressure. Therefore, measurement of blood gas tension should ideally be performed inside the hyperbaric chamber, which requires a dedicated analyzer and a skilled technician.

If these facilities are not available, two approaches to estimation of PaO2 may be taken. The first is to simply draw the arterial blood sample, extrude the bubbles as carefully as possible, decompress the sample, and measure the sample as quickly as possible. Modern blood gas analyzers can produce a PaO2 estimate by extrapolating the calibration curve obtained from lower O2 tensions. Weaver and Howe[173] have demonstrated that O2 tension in blood samples obtained at 3 ATA in a hyperbaric chamber can be measured accurately at 1 ATA if analysis is performed immediately after decompression of the blood sample.

Another approach is to estimate PaO2 based on measurements obtained at 1 ATA. Gilbert and Keighley[64] demonstrated that at 1 ATA, the ratio of arterial-alveolar PO2 (PaO2 /PAO2 , or the a/A ratio) is relatively constant over a wide range of inspired O2 concentrations. We have examined this ratio over a range of inspired O2 concentrations from 0.21 to 1.0 and ambient pressures from 1 to 3 ATA. Arterial blood gases were obtained at 1 ATA, from which the a/A ratio was calculated. Predicted values for PaO2 were then calculated for breathing air or 100% O2 at elevated ambient pressure and correlated with actual measurements of PaO2 inside the chamber. There was reasonable agreement between predicted and actual values ( Fig. 70-11 ) up to PaO2 values of 1700 mm Hg. The following equations may be used to predict PaO2 at pressure.

The alveolar gas equation is required to calculate PAO2 :





where PB and PH2 O are, respectively, the ambient and saturated water vapor pressure and R is the respiratory exchange ratio. If FIO2 = 0.2, R = 0.8, and body temperature = 37°C, Equation 6 can be simplified to

PAO2 = (PB − 47) · 0.2 − 1.2 · PCO2 (7)

Having calculated PAO2 and measured PaO2 , the a/A ratio can then be obtained at 1 ATA. The predicted PaO2 at increased ambient pressure while breathing 100% O2 may then be derived from Equation 7:

PaO2 (pred) = a/A · [(760 · ATA − 47) − PaCO2 ] (8)

where ATA is the chamber pressure in atmospheres absolute. Although dose-response curves for HBO are not yet available, a reasonable aim is to achieve a PaO2 of 1000 mm Hg or higher for routine chronic therapy and as high a level as possible for the treatment of acute necrotizing infections.

A better monitor of tissue oxygenation may be mixed venous PO2 (Pv̄O2 ), which in the absence of shunts, may be a reasonably accurate estimate of mean tissue PO2 . [174]


2682


Figure 70-11 Measured versus predicted PaO2 at increased ambient pressure. Predicted PaO2 is calculated from room-air arterial blood gases, assuming that the arterial-alveolar PO2 ratio (PaO2 /PAO2 , or a/A ratio) is a constant. Data are shown both for persons with normal lungs (a/A ratio ≥0.75) and for patients with gas exchange abnormalities (a/A ratio <0.75). It is evident that PaO2 predicted in this way is close to the actual measured PaO2 . (From Moon RE, Camporesi EM, Shelton DL: Prediction of arterial PO2 during hyperbaric treatment. In Bove AA, Bachrach AJ, Greenbaum LJ Jr [eds]: Underwater and Hyperbaric Physiology IX. Proceedings of the Ninth International Symposium on Underwater and Hyperbaric Physiology. Bethesda, MD, Undersea and Hyperbaric Medical Society, 1987, p 1127.)

In critically ill patients with PA catheters, routine measurement of Pv̄O2 is a convenient method of estimating the degree to which HBO therapy is delivering additional O2 to tissues. A low value may indicate insufficiently high cardiac output to hyperoxygenate tissues. An example is shown in Table 70-8 .

Normal values for pH and PCO2 under hyperbaric conditions are the same as they are at 1 ATA. This relationship has been shown to be valid up to an ambient pressure of 66 ATA.[175] PCO2 (and hence pH) should not be expected to change significantly in blood samples that are decompressed, as noted earlier. Therefore, values of
TABLE 70-8 -- Hemodynamic and oxygen data in a 77-year-old woman with necrotizing fasciitis

Ambient Pressure (ATA) FIO2 PaO2 (mm Hg) Pv̄O2 (mm Hg) Sv̄O2 (%) Cardiac Output * (L/min) Heart Rate (Beats/min)
A 1   2 L/min 130 33 61 3.7 102
B 2.5 1.0 >1000 45 79 2.6  95
C 2.5 1.0 >1000 56 88 3.6 101
*Note the drop in cardiac output and heart rate with hyperbaric oxygenation (A, B). There was a significant increase in Pv̄O2 , which increased further when cardiac output was raised because of a reduction in afterload with sodium nitroprusside infusion (C). At pressure, Pv̄O2 and Sv̄O2 are significantly higher than usually observed at 1 ATA.





PCO2 and pH measured in a blood gas laboratory at 1 ATA on samples of blood decompressed from a chamber can be relied on for clinical purposes.

Mechanical ventilation in a hyperbaric environment presents a variety of challenges. The ideal requirements for ventilation include small size, no electrical requirement, no use of flammable lubricants, ability to operate on a volume-cycled basis over a wide range of tidal volumes and respiratory rates, minimal modification needed for installation, and ability to provide positive end-expiratory pressure, as well as ventilate in intermittent mandatory ventilation and assist/control modes.[176] Additionally, the ideal gas source to actuate the ventilator should minimize the risk of combustion caused by buildup of static electricity.

As ambient pressure increases, gas density is proportionately raised whereas gas viscosity changes relatively little. Therefore, in regions of turbulent flow (i.e., in the large airways), airway resistance would be expected to increase. Measurements of respiratory conductance (the reciprocal of resistance) during tidal breathing[177] indicate that it varies with gas density according to the following formula:

G = G0 ρκ (9)

where G is lung conductance at gas density ρ, G0 is conductance at gas density 1.1 g/L (1 ATA), and κ is a constant that has been found to have a mean value of -0.39. At 6 ATA this equation predicts that lung conductance would decrease by 50%, which is equivalent to a doubling of pulmonary resistance. The increased airway resistance leads to a decrement in maximum voluntary ventilation. [175] [178] In addition, the higher gas density results in less efficient distribution of ventilation manifested as an increase in physiologic dead space.[175] [179] The effects of these two phenomena include higher airway pressure during mechanical ventilation and an increased ventilatory requirement. If ventilator settings are not adjusted to compensate for the higher dead space, a rise in arterial PCO2 will occur.

Several types of ventilators have been used and tested in hyperbaric chambers. Pressure-cycled devices such as the Bird ventilator have been used with some success because their compactness admirably fulfills the requirement for small size. However, continual adjustment of rate and cycling pressure is necessary with changes in ambient pressure. Systematic evaluation of two Bird ventilator models and a Mark 2 model modified for hyperbaric use


2683
by adding inspiratory and expiratory flow cartridges was performed by Gallagher and colleagues.[180] Two of the Bird ventilators failed at 3 ATA and one at 4 ATA. All the devices tested varied their tidal volumes and ventilatory rates with changes in ambient pressure. Several volume-cycled ventilators have been tested and found to be superior. The pneumatic Emerson functioned satisfactorily up to 6 ATA.[180] The disadvantages are that this ventilator was designed to be used only in control mode and its leather bellows are lubricated with mineral oil, which violates most hyperbaric safety standards. The Penlon Oxford ventilator has been tested to 6 ATA in a compressed air environment and to 31 ATA in heliox.[181] It delivered 800-mL tidal volume at stable rates and expired volume over its entire pressure range. The Monaghan 225 ventilator works well in a hyperbaric environment; it delivers up to 18 L/min at 6 ATA over a range of tidal volumes from 0.2 to 2 L.[176] Its main disadvantage is its tendency to decrease the cycling rate with increasing pressure, which requires some adjustment of the rate during compression, and the fact that its standard mode of operation requires supplying it with compressed O2 rather than air. However, a simple modification allows it to be powered with compressed air ( Fig. 70-12 ). We have used this ventilator for hundreds of hours without malfunction. Another ventilator that has been used in a hyperbaric environment


Figure 70-12 Monaghan 225 ventilator modified for use in a hyperbaric chamber. A gas-mixing system with two ball rotameters has been added to provide the inspired gas, which enters the ventilator through a conduit attached to the air intake. The FIO2 setting has been permanently set at 21% to force the ventilator to use the externally mixed gas as the breathing gas. The FIO2 adjustment knob has been removed. A 5-L reservoir bag is attached to the patient gas circuit. A reservoir relief valve prevents overpressurization of the reservoir bag. A subambient pressure relief valve allows entrainment of room air if insufficient gas is provided to the reservoir bag. (From Moon RE, Bergquist LV, Conklin B, Miller JN: Monaghan 225 ventilator use under hyperbaric conditions. Chest 89:846, 1986.)

is the Siemens 900C.[182] This ventilator is widely used in critical care units and would be an excellent clinical tool in the hyperbaric setting because of its flexibility and compactness. However, tidal volume measured by the ventilator is inaccurate at increased ambient pressure because of the use of screen-type pneumotachographs, which have density-dependent characteristics. Other ventilators used in the hyperbaric environment include the Bennett PR-2,[183] Dräger Hyperlog,[184] LAMA-RCH,[185] Logic-03,[186] Penlon Nuffield 200,[187] Pneumatic Emerson,[180] PneuPAC variant HB,[188] Sechrist 500A,[146] and Omnivent.[146]

Two safety considerations are worthy of mention. First, in any ventilator delivering enriched O2 , there is a potential for O2 buildup within the housing. A fire hazard may exist if electrical components surrounded by a high O2 concentration overheat or spark. Moreover, leakage of O2 from the housing into the chamber may cause rapid O2 buildup in the chamber atmosphere. The effect of this leakage on the chamber's O2 concentration can be minimized in practice by inserting scavenging tubing into the ventilator cowling to remove a large proportion of the leaking O2 through the chamber overboard dump system. Modification of the Monaghan ventilator to compressed air actuation [176] eliminates O2 leakage other than from the patient's expired gas. The other safety consideration is the reliability and safety of electrical components. Large sealed


2684
components such as power transistors, Nixie tube displays, and cathode ray tubes are inherently fragile at high ambient pressure. Integrated circuit chips, light-emitting diode displays, and discrete components such as resistors and capacitors are relatively resistant to changes in ambient pressure. Nevertheless, whenever overheating of a component is a potential risk, purging the device with an inert gas will probably prevent ignition (see later).

Air-filled endotracheal tube cuffs tend to lose volume during compression and re-expand during decompression. Therefore, either continuous adjustment of the air pressure within the cuff manually or the easier measure of filling the cuff with water will obviate these difficulties.

Previous Next