|
The most direct method of measuring O2
uptake (V̇O2
)
and CO2
elimination (V̇CO2
)
is by analysis of inspired (I) and expired (E)
gas concentrations. If the minute volume is measured along with inspired and expired
O2
and CO2
concentrations, V̇CO2
can be calculated by the following equation:
V̇CO2
=
V̇E · FECO2
− V̇I
· FICO2
(26)
In Equation 26, V̇E and V̇I
are the expired and inspired respiratory minute volumes, respectively. FECO2
and FICO2
are the mixed expired and inspired
CO2
concentrations, respectively. Because the quantities of O2
consumed and CO2
eliminated are rarely exactly equal, V̇E
and V̇I differ slightly. Because FICO2
is usually zero, Equation 26 can be simplified to the following form:
V̇CO2
=
V̇E · FECO2
(27)
Similarly, an equation can be written for the calculation of O2
consumption:
V̇O2
=
V̇I · FICO2
− V̇E · FECO2
(28)
In practice, it is difficult to measure minute ventilation with
the necessary degree of accuracy to be able to distinguish small differences between
V̇I and V̇E.
Because the exchange of inert gas (i.e., gas in the breathing mixture that is neither
CO2
nor O2
) generally equals zero, V̇I
can be calculated as follows:
FI inert = 1 − FIO2
(30)
FE inert = 1 − (FEO2
+ FECO2
) (31)
Equation 28 can be rewritten as follows:
V̇O2 measurement using this equation becomes progressively less accurate as the inspired O2 fraction is increased above 50%. At high FIO2 values, because FI inert becomes small, errors in the measurement of FI inert result in correspondingly large errors in O2 . At FIO2 = 1, this equation breaks down completely.
Open-circuit measurement of V̇O2 and V̇CO2 under anesthesia using this method has been described by Viale and coworkers.[182] Self-contained analyzers have been designed.[183] [184] These instruments have proved to be extremely satisfactory for gas exchange monitoring in patients in the critical care environment. One practical problem with systems incorporating zirconium oxide O2 sensor is extremely sensitive to small concentrations of fluorinated anesthetic gases, which may cause severe malfunction. V̇O2 and V̇CO2 can be measured on a breath-by-breath basis by integrating FIO2 - FEO2 with respect to exhaled volume.[185] A method of measurement of V̇O2 by using rapid response temperature and humidity analyzers to make appropriate corrections may allow breath-by-breath monitoring even with high FIO2 values.[186]
An alternative technique (i.e., reversed Fick) is to calculate
the O2
consumption as the product of the arteriovenous O2
content
difference and the cardiac output derived by thermodilution:
V̇O2
=
(CaO2
− Cv̄O2
)
· T (33)
In Equation 33, CaO2
and Cv̄O2
are the arterial and mixed venous O2
contents (mL/L), respectively. T
is cardiac output. Blood O2
content may be calculated using Equation
10; when using Equation 31, the O2
content values calculated using Equation
10 must be multiplied by 10 to obtain the O2
in the correct units.
The reversed Fick technique is reasonably satisfactory for clinical purposes, although it becomes less accurate in the setting of high cardiac output and low arteriovenous O2 content differences. It tends to underestimate the V̇O2 measured directly by 30 to 60 mL/min.[187] [188] [189] This difference has been attributed to pulmonary O2 consumption, which is not measured by the reversed Fick method. Use of continuous thermodilution cardiac output versus intermittent bolus injection may reduce the error.[190]
|