Previous Next

Pacemaker Indications

Indications for permanent pacing are shown in Table 35-3 and are reviewed in detail elsewhere.[15] Classically, pacemakers are used to treat patients with diseases of the sinus node (improper impulse formation) and diseases of the AV node (improper impulse conduction). Pacing
TABLE 35-2 -- Rate modulation (activity) sensors
Approved in the United States * Under Investigation *
Vibration sensor Right ventricular stroke volume
Motion sensor Blood pH
Minute ventilation (bio-impedance sensor) Blood temperature
QT interval (Vitatron only) Mixed venous oxygen sensor
Right ventricular pressure (Biotronik only) Systolic time intervals

Evoked response

Intracardiac impedance
*Of the sensors used to detect exercise in a patient with a cardiac pacemaker, five types are approved in the United States, and many others are under investigation. Some devices have two sensors and can be programmed for cross-checking to prevent increases in heart rate from spurious causes. Minute ventilation sensors are very sensitive to stray electromagnetic interference, and patients have been inappropriately treated for pacemaker-driven tachycardias as a result. There is little perioperative experience with right ventricular pressure sensors. Most pacemaker experts[90] recommend programming rate modulation to the Off position in the perioperative period to prevent confusion between an intrinsic tachycardia and a pacemaker-induced tachycardia.





can be used to reduce the outflow tract obstruction in hypertrophic obstructive cardiomyopathy (HOCM) in both adults and children, since paced ventricular conduction takes place in a left bundle branch pattern (left ventricular septum depolarizes after the other segments, rather than as the early systolic event).[16] [17] Finally, in August, 2001, devices were approved by the U.S. FDA for three-chamber pacing (right atrium, both ventricles) to treat dilated cardiomyopathy (DCM).[18] To accomplish left ventricular pacing, a wire is placed through the right atrium and into the coronary sinus (CS) ( Fig. 35-2 ). Most of the current devices connect the electrodes from the CS lead in parallel with the electrodes on the RV lead. In the setting of ventricular noncapture, this parallel connection can lead to ventricular overcounting with resultant inappropriate antitachycardia therapy for the patient with a DCM pacing defibrillator.[19] [20] Devices in clinical trial (as well as the Reveal Defibrillator from
TABLE 35-3 -- Indications for pacemaker placement
Symptomatic diseases of impulse formation (i.e., sinus node disease) *
Symptomatic diseases of impulse conduction (i.e., atrioventricular nodal disease)
Long QT syndrome
Hypertrophic obstructive cardiomyopathy
Dilated cardiomyopathy (also called cardiac resynchronization therapy)
*Indications for permanent pacing are shown. Most patients with pacemakers fall into the first two categories listed.
†Requires 100% ventricular pacing to be effective. Short atrioventricular delays (120 to 150 msec) are programmed.





1419


Figure 35-2 A defibrillator system with biventricular (BiV) pacemaker capability. Three leads are placed: a conventional, bipolar lead to the right atrium; a quadripolar lead to the right ventricle (RV); and a lead to the coronary sinus (CS). This system is designed to provide "resynchronization therapy" in the setting of a dilated cardiomyopathy with a prolonged QRS (and frequently with a prolonged PR interval). The bipolar lead in the right atrium performs sensing and pacing functions. Likewise, the bipolar tip in the RV performs pacing and sensing functions. The lead in the CS depolarizes the left ventricle (LV). In most BiV devices, both ventricles are depolarized simultaneously, because the electrode on the RV lead are connected in parallel with the electrode on the CS lead. This parallel connection can cause ventricular oversensing (and inappropriate antitachycardia therapy) in an implantable cardioverter-defibrillator (ICD) if the pacemaker fails to depolarize one or both of the ventricular chambers. The presence of "shock" conductors on the RV lead, called coils, in the RV and the superior vena cava (SVC) distinguish a defibrillation system from a conventional pacemaking system. Typically, the SVC shock coil is electrically identical to the defibrillator case, called the can. When the defibrillation circuitry includes the ICD case, it is called an active can configuration.

Guidant Medical and InSynch III from Medtronic) have separate sensing and output controls for the ventricular wires.

Pacing for HOCM and DCM requires careful attention to pacer programming. In order to be effective in these patients, the pacemaker must provide the stimulus for ventricular depolarization, and AV synchrony must be preserved.[21] Pacemaker inhibition or loss of pacing (i.e., from native conduction, atrial irregularity, ventricular irregularity, development of junctional rhythm, or electromagnetic interference) can lead to deteriorating hemodynamics in these patients.

Biventricular pacing might cause inappropriate lengthening of the Q-T interval in susceptible patients, and this lengthening has been reported to be associated with torsades de pointes.[22] As a result of this report, the prudent anesthesiologist should ensure adequate access to rapid defibrillation for the patient with biventricular pacing accomplished without an ICD.

Previous Next