BREATHING SYSTEM
*9. Calibrate the O2
Monitor
- Ensure the monitor reads 21% in room air.
- Verify that the low O2
alarm is enabled and functioning.
- Reinstall the sensor in the circuit, and flush the breathing system with
O2
.
- Verify that monitor now reads greater than 90%.
10. Check Initial Status of Breathing System
- Set the selector switch to Bag mode.
- Check that the breathing circuit is complete, undamaged, and unobstructed.
- Verify that the carbon dioxide absorbent is adequate.
- Install the breathing circuit accessory equipment (e.g., humidifier, PEEP
valve) to be used during the case.
11. Perform a Leak Check of the Breathing System
- Set all gas flows to zero (or minimum).
- Close the APL (pop-off) valve, and occlude the Y-piece.
- Pressurize the breathing system to about 30 cm H2
O with an O2
flush.
- Ensure that pressure remains fixed for at least 10 seconds.
- Open the APL (pop-off) valve, and ensure that the pressure decreases.