SCAVENGING SYSTEM

*8. Adjust and Check the Scavenging System

  1. Ensure proper connections between the scavenging system and both the adjustable pressure limiting (APL) (pop-off) valve and the ventilator's relief valve.
  2. Adjust the waste gas vacuum (if possible).
  3. Fully open the APL valve and occlude the Y-piece.
  4. With minimum O2 flow, allow the scavenger reservoir bag to collapse completely, and verify that the absorber pressure gauge reads about zero.
  5. With the O2 flush activated, allow the scavenger reservoir bag to distend fully, and then verify that absorber pressure gauge reads <10 cm H2 O.


316