LOW-PRESSURE SYSTEM
*4. Check Initial Status of the Low-Pressure System
- Close flow control valves, and turn vaporizers off.
- Check the fill level, and tighten the vaporizers' filler caps.
*5. Perform a Leak Check of the Machine's Low-Pressure System
- Verify that the machine master switch and flow control valves are OFF.
- Attach a suction bulb to the common (fresh) gas outlet.
- Squeeze the bulb repeatedly until fully collapsed.
- Verify bulb stays fully collapsed for at
least 10 seconds.
- Open one vaporizer at a time, and repeat steps c and d above.
- Remove the suction bulb, and reconnect the frésh gas hose.
*6. Turn on the Machine's Master Switch and All Other Necessary
Electrical Equipment.
*7. Test Flow Meters
- Adjust flow of all gases through their full range, checking for smooth
operation of floats and undamaged flow tubes.
- Attempt to create a hypoxic O2
/N2
O mixture, and verify
correct changes in the flow and/or alarms.