|
Capnography, the measurement of CO2 in expired gases, has evolved in the last few years into a commonly used procedure. Whereas a variety of techniques can be used for CO2 measurement (e.g., mass spectrometry, Raman analysis), most capnographs rely on infrared absorption.[169] Use of this technique can reliably and quantitatively provide vital respiratory monitoring information in the operating room and in all critical care areas.
End-tidal PO2 (PETO2 ) may be used as an estimate of alveolar PO2 and therefore PaO2 . Whereas end-tidal CO2 (PETCO2 ) analysis has achieved a high degree of popularity, this has not occurred for PO2 monitoring because of the variable A-a gradient. In normal individuals, this gradient may be less than 10 mm Hg, but in patients with severe V̇A/ mismatching, the gradient may be substantially increased. The A-a gradient is increased at high inspired O2 concentrations, even with normal lungs. PETO2 therefore almost always overestimates PaO2 . For example, the PETO2 of a cadaver being ventilated with 100% O2 would be approximately 700 mm Hg (Pbarometric - PH2 O)! Nevertheless, exhaled O2 analysis can be useful in monitoring the adequacy of nitrogen washout in preparation for induction of general anesthesia, particularly when a period of apnea is expected. Clinicians often select the time of induction of anesthesia at a point when the inspired-expired percentage of O2 has decreased to a plateau (typically less than 10%). An example of a continuous tracing of expired PO2 , demonstrating nitrogen washout, is shown in Figure 36-14 .
According to the gas sampling technique, infrared CO2 monitors are in one of two categories: sidestream monitors, which draw a continuous sample of the gas from the respiratory circuit into the measuring cell, and mainstream monitors, which directly straddle the airway with a reading cell placed at the attachment between respiratory circuit and endotracheal tube or breathing mask. The key difference in use between the two types of capnographs depends on details of practical importance and on the type and duration of the monitoring environment.
Sidestream capnographs depend crucially on a sampling flow that continuously aspirates from the side of the main respiratory gas flow a fixed amount of gas. The rate of gas sampling can usually be adjusted from 50 to 500 mL/min and sometimes up to 2 L/min. This continuous bias flow can be the source of significant methodologic error. If the sampling flow ever exceeds the expired gas flow,
Figure 36-14
Monitoring of expired oxygen (O2
) to monitor
lung nitrogen (N2
) washout. If preoxygenation is desired, its progress
can be assessed by monitoring expired O2
. The top panel shows exhaled
O2
concentration on a compressed time scale while breathing 100% O2
through an anesthesia mask. The next panel shows exhaled CO2
on the same
time scale. Exhaled O2
steadily rises, and the difference between the
inhaled and exhaled O2
(I-e on the display) falls.
The most faithful rendition of the capnograph waveform occurs when the sidestream sampling tubing is connected as close to the patient as possible ( Fig. 36-15 ). Monitoring of end-tidal CO2 in the spontaneously breathing patient whose trachea is not intubated requires some improvisation. Nasotracheal cannulae connected to a sidestream monitor usually provide a usable waveform but frequently become obstructed with saliva or mucus
Figure 36-15
Sidestream sampling port placement. A,
To minimize the effects of breathing circuit dead space, attachment of the sampling
port should be as close to the patient as possible (arrow).
B, Placement of the port as shown (arrow)
can cause artifactual lowering of the end-tidal measurement.
|