Previous Next

WAVEFORM ANALYSIS OF EXPIRED RESPIRATORY GASES

Capnography, the measurement of CO2 in expired gases, has evolved in the last few years into a commonly used procedure. Whereas a variety of techniques can be used for CO2 measurement (e.g., mass spectrometry, Raman analysis), most capnographs rely on infrared absorption.[169] Use of this technique can reliably and quantitatively provide vital respiratory monitoring information in the operating room and in all critical care areas.

End-tidal PO2 (PETO2 ) may be used as an estimate of alveolar PO2 and therefore PaO2 . Whereas end-tidal CO2 (PETCO2 ) analysis has achieved a high degree of popularity, this has not occurred for PO2 monitoring because of the variable A-a gradient. In normal individuals, this gradient may be less than 10 mm Hg, but in patients with severe V̇A/ mismatching, the gradient may be substantially increased. The A-a gradient is increased at high inspired O2 concentrations, even with normal lungs. PETO2 therefore almost always overestimates PaO2 . For example, the PETO2 of a cadaver being ventilated with 100% O2 would be approximately 700 mm Hg (Pbarometric - PH2 O)! Nevertheless, exhaled O2 analysis can be useful in monitoring the adequacy of nitrogen washout in preparation for induction of general anesthesia, particularly when a period of apnea is expected. Clinicians often select the time of induction of anesthesia at a point when the inspired-expired percentage of O2 has decreased to a plateau (typically less than 10%). An example of a continuous tracing of expired PO2 , demonstrating nitrogen washout, is shown in Figure 36-14 .

According to the gas sampling technique, infrared CO2 monitors are in one of two categories: sidestream monitors, which draw a continuous sample of the gas from the respiratory circuit into the measuring cell, and mainstream monitors, which directly straddle the airway with a reading cell placed at the attachment between respiratory circuit and endotracheal tube or breathing mask. The key difference in use between the two types of capnographs depends on details of practical importance and on the type and duration of the monitoring environment.

Sidestream Capnographs

Sidestream capnographs depend crucially on a sampling flow that continuously aspirates from the side of the main respiratory gas flow a fixed amount of gas. The rate of gas sampling can usually be adjusted from 50 to 500 mL/min and sometimes up to 2 L/min. This continuous bias flow can be the source of significant methodologic error. If the sampling flow ever exceeds the expired gas flow,


1456


Figure 36-14 Monitoring of expired oxygen (O2 ) to monitor lung nitrogen (N2 ) washout. If preoxygenation is desired, its progress can be assessed by monitoring expired O2 . The top panel shows exhaled O2 concentration on a compressed time scale while breathing 100% O2 through an anesthesia mask. The next panel shows exhaled CO2 on the same time scale. Exhaled O2 steadily rises, and the difference between the inhaled and exhaled O2 (I-e on the display) falls.

contamination from the fresh gas flow source will occur. The sampling gas pump, flow regulator, sampling system (including the connector to the sampling port), and water trap or water separator constitute multiple sites for gas leak or breakage. Depending on the size and length of the sampling tube and the rate of gas flow, a certain delay in gas detection is introduced (i.e., CO2 flight time), which can amount to several seconds when the sampling rate is low and the sampling dead space is high (e.g., long tubes). After measurement in the gas cell, the sampled gas may be exhausted into the atmosphere or retrieved and reinjected through a second tube into the breathing circuit to restore breathing circuit volume. This variable may be of great importance in closed-circuit and precise measurements of metabolic gas volumes. The analytic core of the instrument, the infrared measuring cell, must be carefully protected so that liquids and particulate matter do not enter it and cause erroneous readings of CO2 because of their high infrared absorbance. The major problem is caused by water vapor, which is invariably present in expired air (at 37°C) with a saturated vapor pressure of 47 mm Hg. This condenses at lower (room) temperature on sampling tube walls. In critical care settings and often in the operating room, the inspired gas is kept warm and humid during long cases. This increases the load on water separation systems applied to capnographs. Water traps and filters have been designed to protect the measuring chamber.

The most faithful rendition of the capnograph waveform occurs when the sidestream sampling tubing is connected as close to the patient as possible ( Fig. 36-15 ). Monitoring of end-tidal CO2 in the spontaneously breathing patient whose trachea is not intubated requires some improvisation. Nasotracheal cannulae connected to a sidestream monitor usually provide a usable waveform but frequently become obstructed with saliva or mucus


Figure 36-15 Sidestream sampling port placement. A, To minimize the effects of breathing circuit dead space, attachment of the sampling port should be as close to the patient as possible (arrow). B, Placement of the port as shown (arrow) can cause artifactual lowering of the end-tidal measurement.

and are uncomfortable. Taping a piece of intravenous tubing close to the nostril can provide an estimate of arterial PCO2 that is adequate for clinical purposes. Alternatively, an intravenous catheter can be threaded into the common lumen of a pair of nasal O2 cannulas such that the tip lies midway between the two nasal prongs. The extension tube normally connected to the O2 source is tied off, and the intravenous catheter is then connected to a sidestream capnometer.[170] A commercially available version allows O2 administration while end-tidal CO2 is continuously monitored (Divided Canula, Salter Labs, Arvin, CA). Another device (Oridion, Needham, MA) samples exhalation from the mouth and nose ( Fig. 36-16 ). Sampling CO2 from a facemask, although adequate for monitoring respiratory rate, produces measured PETCO2 values that are significantly lower than PaCO2 .

Previous Next