Preanesthesia Evaluation and Pacemaker Reprogramming
Preoperative management of the patient with a pacemaker includes
evaluation and optimization of coexisting diseases. ACC guidelines suggest that
cardiac testing (stress tests, echocardiograms) should be dictated by the patient's
underlying diseases, medications, symptomatology, interval from the last testing,
and planned intervention.[3]
No special laboratory tests or radiographs are needed for the
patient with a conventional pacemaker. Chest films rarely depict lead problems,
not all devices have radiographic markings, and a standard chest x-ray can exclude
the generator or its markings. However, a patient with a BiV pacer (or defibrillator)
might need a chest film to document the position of the coronary sinus (CS) lead,
especially if central line placement is planned. Currently, there are no data or
guidelines regarding the placement of a central line in the setting of a coronary
sinus lead, and, since these leads have no fixation, they might be more easily dislodged
than a standard pacemaker (or defibrillator) lead. In fact, in early studies, spontaneous
CS lead dislodgement was found in more than 11% of patients.[27]
[28]
Important features of the preanesthetic device evaluation are
shown in Appendix 1
. Current
NASPE and Medicare guidelines include telephonic evaluation every 4–12 weeks
(depending upon device type and age) and at least one direct evaluation (i.e., a
device interrogation with a programmer) at least once per year.[29]
In a recent abstract, Rozner et al reported significant noncompliance
with these guidelines in 161 consecutive patients, as well as the need to replace
5% of pacemakers preoperatively for battery depletion.[30]
Telephonic evaluation provides some assurance of battery longevity, but it does
not provide rigorous evaluation of the pacemaking system. ACC guidelines now suggest
a timely, preoperative interrogation of every pacemaker.[3]
For current pacemakers, interrogation with a programmer remains
the most reliable method for evaluating battery voltage, battery impedance, lead
performance, and adequacy of current settings. Special attention should be paid
to patients from countries where pacemakers might be reused,[31]
[32]
since battery life (typically 5–7 years,
perhaps longer in some new devices) might not be related to length of implantation
in the current patient. For any device with a battery voltage less than 2.6 volts
or battery impedance (where available) greater than 3000 ohms, the device manufacturer
should be consulted about the possible need for elective replacement of the device
prior to the delivery of an anesthetic or surgery.
Appropriate reprogramming ( Table
35-5
) is the safest way to avoid intraoperative problems, especially if
monopolar "Bovie" electrosurgery will be used. The pacemaker manufacturers stand
ready to assist with this task (see Appendix
2
for company telephone numbers), however, any industry-employed allied
professional must be supervised by an appropriately trained physician.[33]
Hospital guidelines and policy should reflect the need for a physician with privileges
to implant pacing devices to review and validate any pacing changes that are made
to a pacemaker by a company representative.
Reprogramming a pacemaker to asynchronous pacing at a rate greater
than the patient's underlying rate usually ensures that no over- or undersensing
during EMI will take place, protecting the patient. However, there is one case report
involving a Telectronics device which would open its output circuitry (i.e., no pacing
stimuli would be delivered) in the presence of high voltage EMI.[34]
At our institution, we have observed a Medtronic Kappa 400 dual chamber device in
the DOO mode drop its rate during use of ESU, presumably due to high battery current
drain.
TABLE 35-5 -- Situations probably requiring pacemaker reprogramming
Any rate-responsive device
*
(Some problems are well known,[90]
[91]
some problems have been misinterpreted with
potential for patient injury,[37]
[41]
[46]
[50]
and the
U.S. FDA has issued an alert regarding devices with minute ventilation sensors[48]
[see Table 35-6
].) |
Special pacing indication (e.g., hypertrophic obstructive cardiomyopathy,
dilated cardiomyopathy, pediatric patients) |
Pacemaker-dependent patient |
Major procedure in the chest or abdomen |
Rate enhancements that should be disabled |
Special procedures |
Lithotripsy |
Transurethral resection |
Hysteroscopy |
Electroconvulsive therapy |
Succinylcholine use |
Magnetic resonance imaging (usually contraindicated by
device manufacturers); might be possible in some pacemaker patients[92]
|
*In
some situations and for certain patients, a pacemaker should be reprogrammed to avoid
potential patient injury or to prevent a pacemaker rhythm that could be confused
with pacemaker malfunction.
Reprogramming a device will not protect the device from internal
damage or reset caused by electromagnetic interference. Additionally, setting a
device to asynchronous mode causes the pacemaker to ignore premature atrial or ventricular
systoles, which could have the potential to create a malignant rhythm in the patient
with significant structural compromise of the myocardium.[35]
In general, rate responsiveness and other "enhancements" (dynamic
atrial overdrive, hysteresis, sleep rate, AV search, etc.) should be disabled by
programming.[3]
[36]
[37]
Note that for many Guidant and/or CPI devices,
Guidant Medical recommends increasing the pacing voltage to "5 volts or higher" in
any case where the monopolar electrosurgical unit will be used. Few cardiologists
follow this recommendation, but there are reports of threshold changes during both
intrathoracic[38]
and non-chest surgery.[39]
Recently, Levine and colleagues reported that significant noncardiac disease states
could increase pacing threshold.[40]
Special attention must be given to any device with a minute ventilation
(bioimpedance) sensor ( Table 35-6
),
because inappropriate tachycardia has been observed secondary to mechanical ventilation,
[41]
[42]
monopolar
"Bovie" electrosurgery,[41]
[43]
[44]
and connection to an electrocardiographic monitor
with respiratory rate monitoring.[45]
[46]
[47]
[48]
[49]
[50]
Sometimes, this pacemaker driven tachycardia
has led to inappropriate, unsuccessful pharmacologic treatment of the tachycardia.
[42]