    Figure 32-28 
Bedside testing of the pulmonary artery catheter-pressure
transducer system.  With the tip of the catheter held at the level of the transducer
and the heart, a pressure of 0 mm Hg is recorded.  The catheter tip is raised to
create a 30-cm vertical fluid column, and the resulting pressure display should equal
22 mm Hg (equivalent to 30 cm H2
O).  (Redrawn from Mark JB:  Atlas
of Cardiovascular Monitoring.  New York, Churchill Livingstone, 1998, Fig. 3-4.)
    
    
    
    
	    
  | 
    
| 
	 | 
| 
	 | 
| 
	 | 
| 
	 | 
| 
	 | 
| 
	 | 
	
  |