Figure 32-28 Bedside testing of the pulmonary artery catheter-pressure transducer system. With the tip of the catheter held at the level of the transducer and the heart, a pressure of 0 mm Hg is recorded. The catheter tip is raised to create a 30-cm vertical fluid column, and the resulting pressure display should equal 22 mm Hg (equivalent to 30 cm H2 O). (Redrawn from Mark JB: Atlas of Cardiovascular Monitoring. New York, Churchill Livingstone, 1998, Fig. 3-4.)