Figure 49-22
Schematic diagram showing the air bubble detection method
for checking adequacy of the seal of the left endobronchial cuff of a left-sided
double-lumen tube. A, When the left lung is selectively
ventilated or exposed to any desired distending pressure and the left cuff is adequately
sealed, no air will escape around the left cuff and out the open right suction port,
and thus no bubbles will be observed passing through the beaker of water. B,
When the left lung is ventilated or exposed to any desired distending pressure and
the left endobronchial cuff is not adequately sealed, air will escape around the
left cuff and out the open right suction port, and thus air bubbles will be observed
passing through the beaker of water. (From Benumof JL: Anesthesia for Thoracic
Surgery. Philadelphia, WB Saunders, 1987.)