|
The passage of even small amounts of electric current through the body can interfere with normal functioning of muscles and nerves. Microshock refers to the direct application of very small voltages or currents to the heart, often intentionally, by means of external or internal cardiac pacemaker electrodes. Inadvertent microshock, however, is dangerous because it can produce ventricular fibrillation. Deaths from accidental microshock were caused by a faulty endoscope and by faulty dialysis equipment.[9] [10] Before 1980, arterial and central venous pressure transducers were a potential source of microshock because they brought electrodes very close to heparinized saline that is in contact with the intravascular compartment.[35] [36] However, in 1982, miniature disposable pressure transducers appeared that could be powered and operated with low-voltage telephone cables. They are now used exclusively, and their safety with respect to microshock has been demonstrated by years of frequent use without reported complications. Such systems, however, are vulnerable to more subtle dangers, as was reported when a faulty cable connection permitted precision pressure tracings but caused a substantially lower pressure reading.[37]
Electric current is the most appropriate variable to address when setting safety limits. Because investigators have reported that electric current passing through the human heart needs to be at least 50 µA before ventricular fibrillation occurs,[33] [38] [39] the American National Standards Institute (ANSI), on December 2, 1993, set 10 µA as the maximum leakage of current allowable through electrodes or catheters contacting the heart. This amount of current is considerably smaller than the peak electric current generated by an implanted pacemaker during a single pulse (0.1 to 10 mA). Because 2 mA is the LIM warning level, the LIM does not provide protection against microshock hazards.
ECG monitoring electrodes are electrically isolated from power circuits in the monitor by an isolation transformer in the main unit, a second level of isolation. However, an anesthesiologist who wishes to perform intravenous or intracardiac ECG monitoring, as occurs during p-wave guided brachial placement of a central venous pressure catheter, should have qualified personnel check the ECG monitor for current leaks. In some ECG and electroencephalographic (EEG) systems, electrode signals from the patient are translated into optical signals by a battery-powered amplifier having a photodiode output that is
Endovascular procedures performed by interventional neuroradiologists include the placement and activation of electrical coagulation coils inside certain types of intracranial aneurysms. Electricity in such coils causes coagulation more slowly than occurs in conventional electrosurgery. The currents in these unipolar devices have not caused macroshock or microshock. Despite the good safety record for such equipment, anesthesiologists monitoring the care of patients in an interventional neuroradiology suite should be aware of potential electrical hazards that come with new techniques.[40]
|