Previous Next

Operating Principles of Ascending Bellows Ventilators

Contemporary examples of ascending bellows, double-circuit, electronic ventilators include the North American Dräger AV-E and the Datex-Ohmeda 7000, 7800, and 7900 series. A generic ascending bellows ventilator is shown in Figure 9-24 . It may be viewed as a breathing bag (i.e., bellows) located within a clear plastic box. The bellows physically separate the driving-gas circuit from the patient's gas circuit. The driving-gas circuit is located outside the bellows, and the patient's gas circuit is inside the bellows. During the inspiratory phase (see Fig. 9-24A ), the driving gas enters the bellows chamber, causing the pressure within it to increase. This increase in pressure is responsible for two events. First, the ventilator's relief valve closes, preventing anesthetic gas from escaping into the scavenging system. Second, the bellows are compressed, and the anesthetic gas within the bellows is delivered to the patient's lungs. This compression action is analogous to the hand of the anesthesiologist squeezing the breathing bag.[51]

During the expiratory phase (see Fig. 9-24B ), the driving gas exits the bellows' chamber. The pressure within the bellows' chamber and the pilot line decline to zero, causing the mushroom portion of the ventilator's relief valve to open. Gas exhaled by the patient fills the bellows before any scavenging occurs. This happens because a weighted ball similar to those used in ball-type positive end-expiratory pressure (PEEP) valves is incorporated into the base of the ventilator's relief valve. The ball produces 2 to 3 cm H2 O of backpressure; therefore scavenging occurs only after the bellows fill completely and the pressure inside the bellows exceeds this pressure threshold. This design causes all ascending bellows ventilators to produce 2 to 3 cm H2 O of PEEP within the breathing circuit. Scavenging occurs only during the expiratory phase because the ventilator relief valve is open only during expiration.[51]

Gas flow from the anesthesia machine into the breathing circuit is continuous and independent of ventilator activity. During the inspiratory phase of mechanical


300
ventilation, the ventilator's relief valve is closed, and the breathing system's APL valve (pop-off valve) is closed or out of circuit. The patient receives a volume from the bellows and flow meters during the inspiratory phase. Factors that influence the correlation between set tidal volume and exhaled tidal volume include the flow meter's settings, the inspiratory time, the compliance of the breathing circuit, external leakage, and the location of the tidal volume sensor.[15] [16] [139] [140] Usually, the volume gained from the flow meters during inspiration is counteracted by the volume lost to the breathing circuit compliance. The set tidal volume generally approximates the exhaled tidal volume. However, oxygen flushing during the inspiratory phase can result in barotrauma because excess volume cannot be vented.[51]

Previous Next