Pacemaker and Cardioverter-Defibrillator Implantation
(also see Chapter
35
)
Permanent pacemakers and permanent ICDs are frequently implanted
in the cardiac catheterization laboratory. Implantation of both devices involves
placement of transvenous leads into the cardiac chambers (usually the right atrium
or ventricle or both) with tunneling of the leads to a subcutaneous pocket in which
the device is placed. The discomfort of the procedure can be reduced with local
anesthesia, and many of these procedures are performed with sedation/analgesia.
General anesthesia is necessary during testing of the permanent cardioverter-defibrillator.
When the ICD is tested, ventricular tachycardia or fibrillation is induced, and
the ability of the device to sense the arrhythmia and deliver appropriate cardioversion-defibrillation
energy is confirmed. Many patients undergoing ICD placement have poor ventricular
function. Return of hemodynamic variables to baseline after cardioversion-defibrillation
must be closely monitored. It is our practice to place radial arterial catheters
for continuous blood pressure monitoring in patients with a history of hemodynamic
instability, as well as in all patients with a documented left ventricular ejection
fraction less than 20%.
|