Low-Flow Anesthetic Delivery
Low-flow anesthetic delivery produces more stability than closed-circuit
delivery. Low-flow delivery can provide most of the advantages of closed systems,
decreasing instability, while retaining considerable advantages over open systems
in economy, maintenance of humidification and temperature, and atmospheric pollution.
Relative to closed-circuit delivery, low-flow delivery produces a constancy of oxygen
and anesthetic levels and provides greater elimination of carbon monoxide and toxic
anesthetic breakdown products.
In a low-flow delivery system, two factors control the relationship
between the concentration delivered from a vaporizer (FD)
and that in the alveoli (FA), a relationship that
may best be described by the ratio (FD/FA)
of the two variables. First, we have already seen that uptake governs this ratio
in a closed system (see Fig. 5-17A
).
FD/FA is higher
for more soluble anesthetics, and regardless of solubility, the ratio is highest
early in anesthetic administration and decreases rapidly in the first 5 to 10 minutes
of anesthesia (as uptake by the VRG of tissues decreases to the point of near-equilibration)
and more slowly there-after (as uptake by MG, FG, and the fourth compartment decreases).
Second, inflow rate also governs FD/FA.
The relationship is inverse: The higher the inflow rate, the lower is the ratio
(compare Fig. 5-17B to D
with Fig. 5-17A
and notice
the large scaling differences associated with increasing inflow rate). An increase
in inflow rate decreases FD/FA
by decreasing rebreathing. Rebreathing is important because uptake of anesthetic
depletes the anesthetic concentration in rebreathed gases, and the concentration
in FD must be sufficient to compensate for this depletion.
The higher the inflow rate, the less compensation is required because rebreathing
is reduced.
However, increases in inflow rate do not proportionally decrease
FD/FA (see Fig.
5-17A to D
). The greatest reduction in FD/FA
comes with only modest increases in inflow rate. A large decrease occurs when the
inflow rate is changed from that needed for a closed circuit to an inflow rate of
1 L/min, whereas only a small decrease occurs when the inflow rate is increased from
2 to 4 L/min. After the inflow rate exceeds minute ventilation (i.e., a non-rebreathing
system exists), further increases in inflow rate have no effect on FD/FA,
and FD/FA is the
same as the ratio of FI to FD
(i.e., FD/FA = FI/FA).
I use the term anesthetic tether
as a metaphor for the FD/FA
ratio.[82]
A large ratio equates to a long tether,
one permitting considerable freedom or variability in the alveolar concentration.
With a long tether, changes in uptake caused by changes in physiologic variables
(e.g., an increase in cardiac output consequent to surgical stimulation) can appreciably
alter the alveolar concentration and the level of anesthesia. In the example just
cited, there is positive feedback because, by increasing uptake, surgical stimulation
decreases the alveolar concentration and thereby increases the perception of that
stimulation. Most anesthetists prefer a short anesthetic tether because a short
tether provides a tighter control over the level of anesthesia. The use of less
soluble anesthetics and higher inflow rates shortens the tether.
Another benefit to a short tether accrues to the anesthetist who
does not use an agent-specific analyzer. In the absence of such an analyzer, some
anesthetists rely on the dial setting of the vaporizer to indicate the concentration
of anesthetic in the patient's lungs (i.e., assume that FD
equals FA). The vaporizer setting may correlate
poorly with the concentration in the lungs under three circumstances: early in anesthesia
for all agents, later in anesthesia for closed circuits or very low inflow rates,
and later in anesthesia for more soluble agents, such as isoflurane, even at higher
inflow rates. The vaporizer setting may correlate well with poorly soluble anesthetics
such as sevoflurane and desflurane 30 to 60 minutes after the inception of anesthetic
administration. At this time, the delivered concentration from the vaporizer may
be less than 20% greater than that in the alveoli (i.e., the FD/FA
ratio is 1.2), even at inflow rates of 1 to 2 L/min (see Fig.
5-17B and C
).
Economic concerns increasingly dictate the practice of anesthesia,
and anesthesiologists may wish to appreciate the differences in anesthetic consumption
as a function of the choice of inflow rate, the duration of anesthesia, and the choice
of anesthetic. The reports by Yasuda and associates[11]
[12]
provide constants that can be used to estimate
the uptake of commonly available potent inhaled anesthetics. By using the gas laws
and published values for specific gravities, the values for uptake of vapor may be
converted to milliliters of liquid taken up. Combining this information with a knowledge
of the function of circuit rebreathing systems[74]
allows an estimate of the amounts of liquid in milliliters that must be delivered
at various inflow rates to provide a constant alveolar concentration equal to the
MAC[83]
( Table
5-4
). The relative costs of anesthesia may be estimated by applying the
price of the anesthetic of interest to the number of milliliters needed to sustain
anesthesia.
If economy and a low FD/FA
ratio are desirable, the aforementioned considerations suggest that a good compromise
is the use of a low-flow delivery system after an initial period of higher flows.
Higher flows (4 to 6 L/min) may be applied early in anesthesia (i.e., at the times
of highest uptake) and then decreased progressively as uptake decreases. Flows of
2 to 4 L/min might be given for the period from 5 to 15 minutes after inducing anesthesia,
and flows of 1 L/min may be given thereafter. If the average inflow rate were 1
L/min, 1 hour of anesthesia with the four potent anesthetics listed in Table
5-4
would require administration of 6.5 (halothane) to 26.1 (desflurane)
mL of liquid. This fourfold range of values is smaller than the eightfold range
of potency (MAC) values because the amount of anesthetic delivered must account for
more than potency. The amount delivered also must compensate for uptake and losses
of anesthetic through the overflow valve. The relatively smaller uptake and
losses of the less soluble desflurane and sevoflurane are what account for the reduction
from eightfold to fourfold. An even smaller range is found at lower inflow rates,
decreasing to about twofold for a closed circuit. However, such flows should not
be used with sevoflurane because of the greater concentrations of compound A that
result.