TRANSCUTANEOUS GAS TENSION MEASUREMENT
In areas of the skin where local blood flow exceeds the amount
required for local O2
consumption, capillary PO2
may approximate PaO2
. This may be particularly
true if the local area is warmed. This principle has been exploited by manufacturers
of transcutaneous PO2
measuring instruments.
These devices usually consist of a small electrode that is attached with adhesive
to the skin. The skin is locally warmed to 40°C or 41°C. O2
from capillaries diffuses through the intact skin into a Clark-type electrode that
measures PO2
directly. This value usually
correlates well with PaO2
.[45]
However, in the presence of peripheral vasoconstriction[46]
or with thick (adult) skin, the measurement may be erroneous. A reduction in cardiac
output tends to result in an artifactual decrease in transcutaneous PO2
because of the ensuing cutaneous hypoxia. Peripheral vascular disease also reduces
the transcutaneous PO2
value.[47]
[48]
Transcutaneous gas monitoring is particularly useful in infants,
in whom local skin blood flow tends to be high and in whom repeated withdrawals of
arterial blood may cause anemia. These instruments require frequent calibration.
The time constant of measurement is relatively long, and a sudden decrease in PaO2
may not be detectable quickly enough to allow a timely therapeutic response. Skin
burns have sometimes resulted from prolonged application of these devices.
Transcutaneous PCO2
monitoring devices are also available.[49]
Because
changes in transcutaneous PCO2
are slow
to detect apnea or airway disconnects, measurement of these values by monitors is
less widely used than monitoring of end-tidal CO2
(see "Expired Gas Analysis").