APPENDIX: Anesthesia Apparatus Checkout Recommendations,
1993
This checkout list or a reasonable equivalent should be conducted
before administration of anesthesia. These recommendations are valid only for an
anesthesia system that conforms to current and relevant standards and that includes
an ascending bellows ventilator and at least the following monitors: capnograph,
pulse oximeter, oxygen analyzer, respiratory volume monitor (i.e., spirometer), and
breathing system pressure monitor with high-pressure and low-pressure alarms. Users
are encouraged to modify these guidelines to accommodate differences in equipment
design and variations in local clinical practice. Such local modifications should
have appropriate peer review. Users should refer to the operator's manual for the
manufacturer's specific procedures and precautions, especially the manufacturer's
low-pressure leak test (step 5).
EMERGENCY VENTILATION EQUIPMENT
*1. Verify Backup Ventilation Equipment Is Available and Functioning