KEY POINTS
- In the event of a pipeline crossover, two actions must be taken. The backup
oxygen cylinder must be on, and the wall supply sources must be disconnected.
- Fail-safe valves and proportioning systems help minimize delivery of a
hypoxic mixture, but they are not foolproof. Delivery of a hypoxic mixture can result
from the wrong supply gas, a defective or broken safety device, leaks downstream
from the safety devices, inert gas administration, and dilution of the inspired oxygen
concentration by high concentrations of inhaled anesthetics.
- Because of desflurane's low boiling point and high vapor pressure, controlled
vaporization of desflurane requires special, sophisticated vaporizers such as the
Datex-Ohmeda Tec 6 and the Aladin Cassette Vaporizer.
- Misfilling an empty variable-bypass vaporizer with desflurane could be
catastrophic, resulting in delivery of a hypoxic mixture and a massive overdose of
inhaled desflurane anesthetic.
- Inhaled anesthetics can interact with carbon dioxide absorbents and produce
toxic compounds. During sevoflurane anesthesia, compound A can be formed, particularly
at low rates of fresh gas flow, and during desflurane anesthesia, carbon monoxide
can be produced, particularly with desiccated absorbents.
- Anesthesia ventilators with ascending bellows (i.e., bellows that ascend
during the expiratory phase) are safer than those with descending bellows because
disconnections readily manifest with ascending bellows.
- When using anesthesia ventilators with ascending bellows, during the inspiratory
phase, fresh gas flow and oxygen flushing contribute to the patient's tidal volume
because the ventilator's relief valve is closed. Oxygen flushing during the inspiratory
phase can cause barotrauma, particularly in pediatric patients. The anesthesiologist
must never activate the oxygen flush during the inspiratory phase of mechanical ventilation.
- New ventilators that use FGD technology virtually eliminate the possibility
of barotrauma by oxygen flushing during the inspiratory phase because fresh gas flow
and oxygen flush flow are diverted to the reservoir breathing bag. However, if the
breathing bag has a leak or is absent, the patient may become aware while under anesthesia,
and a lower than expected oxygen concentration may be delivered because of entrainment
of room air.
- With newer Ohmeda anesthetic ventilators such as the 7100 and 7900 SmartVent,
the gas from the patient and the drive gas are scavenged, resulting in substantially
increased volumes of scavenged gas. The scavenging systems must be set appropriately
to accommodate the increased volume, or pollution of the operating room environment
can result.
- The low-pressure circuit is the vulnerable area of the anesthesia machine
because it is most subject to breakage and leaks. The low-pressure circuit is located
downstream from all anesthesia machine safety features except the oxygen analyzer,
and it is the portion of the machine that is missed if an inappropriate leak test
for a low-pressure circuit is performed.
- It is mandatory to check the low-pressure circuit for leaks before an anesthetic
is delivered because leaks in the circuit can cause delivery of a hypoxic mixture
or cause the patient to become aware during anesthesia, or both.
- Because most Ohmeda anesthesia machines have a one-way check valve in the
low-pressure circuit, a negative-pressure leak test is required to detect leaks.
A positive-pressure leak test cannot detect leaks in the low-pressure circuit of
most Datex-Ohmeda products.
- Internal vaporizer leaks can be detected only with the vaporizer turned
on.
- Before an anesthetic is administered, the circle system must be checked
for leaks and for flow. To test for leaks, the circle system is pressurized to 30
cm H2
O, and the circle system's airway pressure gauge is observed (i.e.,
static test). To check for appropriate flow to rule out obstructions and faulty
valves, the ventilator and a test lung (e.g., breathing bag) are used (i.e., dynamic
test).
- Newer anesthesia workstation self-tests do not detect internal vaporizer
leaks unless each vaporizer is individually turned on during the self-test.
|