|
Oxygen and nitrous oxide supply sources existed as independent entities in older models of anesthesia machines, and they were not pneumatically or mechanically interfaced. Abrupt or insidious oxygen pressure failure could lead to the delivery of a hypoxic mixture. The 2000 ASTM F1850-00 standard states, "The anesthesia gas supply device shall be designed so that whenever oxygen supply pressure is reduced to below the manufacturer specified minimum, the delivered oxygen concentration shall not decrease below 19% at the common gas outlet."[7] Contemporary anesthesia machines have a number of safety devices that act together in a cascade manner to minimize the risk of hypoxia as oxygen pressure decreases. Several of these devices are described in the following sections.
Many older anesthesia machines have a pneumatic alarm device that sounds a warning when the oxygen supply pressure decreases to a predetermined threshold value such as 30 psig. The 2000 ASTM F1850-00 standard mandates that medium priority alarm shall be activated within 5 seconds when the oxygen pressure decreases below a manufacturer-specified threshold pressure.[7] Electronic alarm devices are used to meet this guideline.
A fail-safe valve is present in the gas line supplying each of the flow meters except that for oxygen. Controlled by oxygen supply pressure, the valve shuts off or proportionally decreases the supply pressure of all other gases (e.g., nitrous oxide, air, carbon dioxide, helium, nitrogen) as the oxygen supply pressure decreases. Unfortunately, the misnomer "fail-safe" has led to the misconception that the device prevents administration of a hypoxic mixture. This is not the case. Machines that are not equipped with a flow proportioning system (see "Proportioning Systems") can deliver a hypoxic mixture under normal working conditions. The oxygen flow control valve can be closed intentionally or accidentally. Normal oxygen pressure keeps other gas lines open so that a hypoxic mixture can result.[8] [9]
Datex-Ohmeda machines are equipped with a fail-safe valve known as the pressure sensor's shut-off valve ( Fig. 9-4 ). The valve operates on a threshold principle, and it is open or closed. Oxygen supply pressure opens the valve, and the valve's return spring closes the valve. Figure 9-4 shows a nitrous oxide pressure sensor's shut-off valve with a threshold pressure of 20 psig.[11] [12] [14] [15] [16] In Figure 9-4A , an oxygen supply pressure greater than 20 psig is exerted on the mobile diaphragm. This pressure moves the piston and pin upward, and the valve opens. Nitrous oxide flows freely to the nitrous oxide flow control valve. In Figure 9-4B , the oxygen supply pressure is less than 20 psig, and the force of the valve's return spring completely closes the valve.[9] Nitroux oxide flow stops at the closed fail-safe valve, and it does not advance to the nitrous oxide flow control valve.
Dräger Medical uses a fail-safe valve known as the oxygen failure protection device (OFPD) to interface the oxygen pressure with that of other gases, such as nitrous oxide or other inert gases.[19] [20] [21] It differs from Datex-Ohmeda's oxygen pressure sensor's shut-off valve because the OFPD is based on a proportioning principle rather than a threshold principle. The pressure of all gases controlled by the OFPD decreases proportionally with the oxygen pressure. The OFPD consists of a seat-nozzle assembly
Figure 9-4
Shut-off valve of the pressure sensor. A,
The valve is open because the oxygen supply pressure is greater than the threshold
value of 20 psig. B, The valve is closed because
of inadequate oxygen pressure. (Adapted from Bowie E, Huffman LM: The Anesthesia
Machine: Essentials for Understanding. Madison, WI, Ohmeda, a Division of BOC Health
Care, 1985.)
Most contemporary Datex-Ohmeda workstations have a second-stage pressure regulator for oxygen that is set at a specific value between 12 and 19 psig. [11] [12] [13] [14] [15] [16] The oxygen flow meter's output is constant when the oxygen supply pressure exceeds the threshold value. The Datex-Ohmeda pressure sensor's shut-off valves are set at a higher threshold value (20 to 30 psig). This ensures that oxygen is the last gas flow to decrease if oxygen pressure fails.
|