Previous Next

Generic Anesthesia Machine

A diagram of a generic two-gas anesthesia machine is shown in Figure 9-3 . The pressures within the anesthesia machine can be divided into three circuits: a high-pressure, an intermediate-pressure, and a low-pressure circuit (see Fig. 9-3 ). The high-pressure circuit is confined to the cylinders and the cylinders' primary pressure regulators. For oxygen, the pressure range of the high-pressure circuit extends from a high of 2200 pounds per square inch gauge (psig) to 45 psig, which is the regulated cylinder pressure. For nitrous oxide in the high-pressure circuit, pressures range from a high of 750 psig in the cylinder to a low of 45 psig. The intermediate-pressure circuit begins at the regulated cylinder supply sources at 45 psig, includes the pipeline sources at 50 to 55 psig, and extends to the flow control valves. Depending on the manufacturer and specific machine design, second-stage pressure regulators may be used to decrease the pipeline supply pressures to the flow control valves to even lower pressures, such as 14 or 26 psig within the intermediate-pressure circuit. [8] [9] [10] [11] [12] [13] [14] [15] [16] The low-pressure circuit extends from the flow control valves to the common gas outlet. The low-pressure circuit includes the flow tubes, the vaporizers, and a one-way check valve on most Datex-Ohmeda machines.[8]

Both oxygen and nitrous oxide have two supply sources: a pipeline supply source and a cylinder supply source. The pipeline supply source is the primary gas source for the anesthesia machine. The hospital piping system provides gases to the machine at approximately 50 psig, which is the normal working pressure of most machines.


275


Figure 9-3 Diagram of a generic two-gas anesthesia machine. (Modified from Check-Out, a Guide for Preoperative Inspection of an Anesthesia Machine. Park Ridge, IL, American Society of Anesthesiologists, 1987.)

The cylinder supply source serves as a backup if the pipeline fails. The oxygen cylinder source is regulated from 2200 to approximately 45 psig, and the nitrous oxide cylinder source is regulated from 745 to approximately 45 psig.[8] [9] [10] [11] [12] [13] [14] [15] [16] [17] [18] [19] [20] [21] [22] [23] [24] [25] [26] [27] [28]

A safety device traditionally referred to as the fail-safe valve is located downstream from the nitrous oxide supply source. It serves as an interface between the oxygen and nitrous oxide supply sources. This valve shuts off or proportionally decreases the supply of nitrous oxide (and other gases) if the oxygen supply pressure decreases. Contemporary machines have an alarm device to monitor the oxygen supply pressure. A high-priority alarm is actuated when the oxygen supply pressure declines to a predetermined threshold, such as 30 psig.[8] [9] [19] [28]

Most Datex-Ohmeda machines have a second-stage oxygen regulator located downstream from the oxygen supply source in the intermediate-pressure circuit. It is adjusted to a precise pressure level, such as 14 psig.[9] [10] [11] [12] [13] [14] [15] [16] This regulator supplies a constant pressure to the oxygen flow control valve, regardless of fluctuating oxygen pipeline pressures. For example, the flow from the oxygen flow control valve remains constant if the oxygen supply pressure is greater than 14 psig.

The flow control valves represent an important landmark because they separate the intermediate-pressure circuit from the low-pressure circuit. The low-pressure circuit is the part of the machine that is downstream from the flow control valves. The operator regulates flow entering the low-pressure circuit by adjusting the flow control valves. The oxygen and nitrous oxide flow control valves are linked mechanically or pneumatically by a proportioning system to help prevent delivery of a hypoxic mixture. The flow travels through a common manifold and may be directed to a calibrated vaporizer. Precise amounts of inhaled anesthetic can be added, depending on the setting of the vaporizer control dial. The total fresh gas flow then travels toward the common gas outlet.[8] [9]

Many Datex-Ohmeda anesthesia machines have a one-way check valve between the vaporizers and the common gas outlet.[9] [10] [11] [12] [13] [14] [17] [18] Its purpose is to prevent backflow into the vaporizer during positive-pressure ventilation, thereby minimizing the effects of downstream, intermittent pressure fluctuations on inhaled anesthetic concentration (see "Intermittent Backpressure"). The presence or absence of this check valve profoundly influences which preoperative leak test is indicated (see "Checking Anesthesia Machines"). The oxygen flush connection joins the mixed-gas pipeline between the oneway check valve (when present) and the machine outlet. When the oxygen flush valve is activated, the pipeline oxygen pressure has a "straight shot" to the common gas outlet.[8] [9]

Previous Next