Previous Next

Implantable Cardioverter-Defibrillator Magnets

Like pacemakers, magnet behavior in ICDs can be altered by programming. Most devices will suspend tachydysrhythmia detection (and therefore therapy) when a magnet is appropriately placed to activate the reed switch. Some devices from Angeion, CPI, Pacesetter (St. Jude Medical) or Ventritex can be programmed to ignore magnet placement.

Depending on programming, antitachycardia therapy in some Guidant and/or CPI devices can be permanently


*The capacitor in an ICD "deforms" during inactivity, leading to increased time needed to charge the capacitor. To mitigate the effects of deformation, all ICDs perform a nontherapeutic charging of their capacitor (called "reforming") at a programmable periodic interval (usually one to six months).

1429

TABLE 35-8 -- Indications for implantable cardioverter-defibrillator placement
Ventricular tachycardia
Ventricular fibrillation
Brugada syndrome (i.e., right bundle branch block, ST-segment elevation in V1 -V3 )
Arrhythmogenic right ventricular dysplasia
Long QT syndrome
Hypertrophic cardiomyopathy
Prophylactic use in the post-myocardial infarction patient who has an ejection fraction less than 30% (i.e., no electrophysiologic testing is needed to justify placement)[72]
Prophylactic use in any patient with a cardiomyopathy from any cause who has an ejection fraction less than 35% *
*Sudden cardiac death—heart failure trial. Presented at the American College of Cardiology Meeting, March 2004.




disabled by magnet placement for 30 seconds, and, as noted above, some patients have been discovered with their ICD antitachycardia therapy unintentionally disabled. [
2] In Guidant and CPI devices, if magnet mode is enabled and the ICD is enabled for antitachycardia therapy, the ICD will emit beeps synchronized to R-waves, signifying adequate placement of magnet and suspension of tachyarrhythmia detection (thereby disabling therapy). If the ICD emits a constant tone, then antitachycardia therapy has been programmed to "off." Depending upon programming, ICDs from Guidant and CPI can toggle between activated and inactivated states with 30 seconds of magnet application. To re-enable therapy, the magnet must be removed, then replaced until the constant tone reverts to beeps synchronized to R-waves. Subsequent removal of the magnet then returns the antitachycardia therapy to the enabled state. Any Guidant or CPI device that emits either beeps or constant tones while a magnet is placed will have its antitachycardia therapy disabled during the magnet session.

To ensure correct magnet placement on their devices, Medtronic markets a device called the "Smart Magnet." This battery-powered instrument contains a magnet and a radio-frequency (RF) receiver. All Medtronic generators broadcast RF destined for the programmer upon magnet placement, and the Smart Magnet detects this radiofrequency transmission and illuminates a light to indicate that RF is being received. Medtronic recommends that the Smart Magnet be taped onto the patient. During the occurrence of EMI, the transmission from the ICD to the Smart Magnet is interrupted, and the ICD "found" light often turns off. In this setting, however, the ICD remains disabled, since it is the physical presence of the magnet, rather than the radiofrequency communication, that disables the antitachycardia therapy.

In general, magnets will not affect antibradycardia pacing mode or rate (except ELA [rate change] and Telectronics Guardian 4202/4203 [disabled]). Intermedics devices transiently change pacing rate (VVI mode) to reflect battery voltage. Interrogating the device and calling the manufacturer remain the most reliable method for determining magnet response.

Previous Next